Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

CCL17 ELISA Kit

CCL17 Reactivity: Human Colorimetric Sandwich ELISA 5-2500 pg/mL Cell Culture Supernatant, Plasma, Serum
Catalog No. ABIN625091
  • Target See all CCL17 ELISA Kits
    CCL17 (Chemokine (C-C Motif) Ligand 17 (CCL17))
    Reactivity
    • 6
    • 5
    • 3
    • 3
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    Human
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Detection Range
    5-2500 pg/mL
    Minimum Detection Limit
    5 pg/mL
    Application
    ELISA
    Purpose
    Human TARC (CCL17) ELISA Kit for cell culture supernatants, plasma, and serum samples.
    Sample Type
    Plasma, Cell Culture Supernatant, Serum
    Analytical Method
    Quantitative
    Specificity
    This ELISA kit shows no cross-reactivity with the following cytokines tested: human Angiogenin, BDNF, BLC, ENA-78, FGF- 4, IL-1 alpha, IL-1 beta, IL-2, IL-3, IL-4, IL-5, IL-7, IL-8, IL-9, IL-10, IL-11, IL-12 p70, IL-12 p40, IL-13, IL-15, I-309, IP-10, G-CSF, GM-CSF, IFN-gamma, Leptin, MCP-1, MCP-2, MCP-3, MDC, MIP-1 alpha, MIP-1 beta, MIP-1 delta, PARC, PDGF, RANTES, SCF, TGF-beta, TIMP-1, TIMP-2, TNF-alpha, TNF-beta, TPO, VEGF.
    Sensitivity
    5 pg/mL
    Characteristics
    • Strip plates and additional reagents allow for use in multiple experiments
    • Quantitative protein detection
    • Establishes normal range
    • The best products for confirmation of antibody array data
    Components
    • Pre-Coated 96-well Strip Microplate
    • Wash Buffer
    • Stop Solution
    • Assay Diluent(s)
    • Lyophilized Standard
    • Biotinylated Detection Antibody
    • Streptavidin-Conjugated HRP
    • TMB One-Step Substrate
    Material not included
    • Distilled or deionized water
    • Precision pipettes to deliver 2 μL to 1 μL volumes
    • Adjustable 1-25 μL pipettes for reagent preparation
    • 100 μL and 1 liter graduated cylinders
    • Tubes to prepare standard and sample dilutions
    • Absorbent paper
    • Microplate reader capable of measuring absorbance at 450nm
    • Log-log graph paper or computer and software for ELISA data analysis
    Featured
    Discover our best selling CCL17 ELISA Kit
    Top Product
    Discover our top product CCL17 ELISA Kit
  • Application Notes
    Recommended Dilution for serum and plasma samples2 fold
    Sample Volume
    100 μL
    Plate
    Pre-coated
    Protocol
    1. Prepare all reagents, samples and standards as instructed in the manual.
    2. Add 100 μL of standard or sample to each well.
    3. Incubate 2.5 h at RT or O/N at 4 °C.
    4. Add 100 μL of prepared biotin antibody to each well.
    5. Incubate 1 h at RT.
    6. Add 100 μL of prepared Streptavidin solution to each well.
    7. Incubate 45 min at RT.
    8. Add 100 μL of TMB One-Step Substrate Reagent to each well.
    9. Incubate 30 min at RT.
    10. Add 50 μL of Stop Solution to each well.
    11. Read at 450 nm immediately.
    Reagent Preparation
    1. Bring all reagents and samples to room temperature (18 - 25°C) before use. 2. Sample dilution: Assay Diluent C (Item L) is used for dilution of serum/plasma/cell culture supernate/urine. 3. Assay Diluent B should be diluted 5-fold with deionized or distilled water before use. 4. Preparation of standard: Briefly spin the vial of Item C and then add 400 µl Assay Diluent C into Item C vial to prepare a 50 ng/ml standard. Dissolve the powder thoroughly by a gentle mix. Add 30 µl TARC standard (50 ng/ml) from the vial of Item C, into a tube with 570 µl Assay Diluent C to prepare a 2,500 pg/ml standard solution. Pipette 300 µl Assay Diluent C into each tube. Use the 2,500 pg/ml standard solution to produce a dilution series (shown below). Mix each tube thoroughly before the next transfer. Gently vortex to mix. Assay Diluent C serves as the zero standard (0 pg/ml). 5. If the Wash Concentrate (20x) (Item B) contains visible crystals, warm to room temperature and mix gently until dissolved. Dilute 20 ml of Wash Buffer Concentrate into deionized or distilled water to yield 400 ml of 1x Wash Buffer. 6. Briefly spin the Detection Antibody vial (Item F) before use. Add 100 µl of 1x Assay Diluent B into the vial to prepare a detection antibody concentrate. Pipette up and down to mix gently (the concentrate can be stored at 4°C for 5 days). The detection antibody concentrate should be diluted 80-fold with 1x Assay Diluent B and used in step 4 of Part VI Assay Procedure. 7. Briefly spin the HRP-Streptavidin concentrate vial (Item G) and pipette up and down to mix gently before use. HRP-Streptavidin concentrate should be diluted 20,000-fold with 1x Assay Diluent B. For example: Briefly spin the vial (Item G) and pipette up and down to mix gently . Add 2 µl of HRP-Streptavidin concentrate into a tube with 198 µl 1x Assay Diluent B to prepare a 100-fold diluted HRP- Streptavidin solution (do not store the diluted solution for next day use). Mix through and then pipette 60 µl of prepared 100-fold diluted solution into a tube with 12 ml 1x Assay Diluent B to prepare a final 20,000 fold diluted HRP-Streptavidin solution.
    Assay Procedure
    1. Bring all reagents and samples to room temperature (18 - 25°C) before use. It is recommended that all standards and samples be run at least in duplicate. 2. Add 100 µl of each standard (see Reagent Preparation step 2) and sample into appropriate wells. Cover well and incubate for 2.5 hours at room temperature or over night at 4°C with gentle shaking. 3. Discard the solution and wash 4 times with 1x Wash Solution. Wash by filling each well with Wash Buffer (300 µl) using a multi-channel Pipette or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels. 4. Add 100 µl of 1x prepared biotinylated antibody (Reagent Preparation step 6) to each well. Incubate for 1 hour at room temperature with gentle shaking. 5. Discard the solution. Repeat the wash as in step 3. 6. Add 100 µl of prepared Streptavidin solution (see Reagent Preparation step 7) to each well. Incubate for 45 minutes at room temperature with gentle shaking. 7. Discard the solution. Repeat the wash as in step 3. 8. Add 100 µl of TMB One-Step Substrate Reagent (Item H) to each well. Incubate for 30 minutes at room temperature in the dark with gentle shaking. 9. Add 50 µl of Stop Solution (Item I) to each well. Read at 450 nm immediately.
    Calculation of Results

    Calculate the mean absorbance for each set of duplicate standards, controls and samples, and subtract the average zero standard optical density. Plot the standard curve on log-log graph paper or using Sigma plot software, with standard concentration on the x-axis and absorbance on the y-axis. Draw the best-fit straight line through the standard points.

    Restrictions
    For Research Use only
  • Storage
    -20 °C
    Storage Comment
    The entire kit may be stored at -20°C for up to 1 year from the date of shipment. Avoid repeated freeze-thaw cycles. The kit may be stored at 4°C for up to 6 months. For extended storage, it is recommended to store at -80°C.
    Expiry Date
    6 months
  • Su, Zhao, He, Huang, Liu, Chen, Chen, Liao, Cui, Zeng, Yao, Su, Liu, Jiang, Song: "miR-142-5p and miR-130a-3p are regulated by IL-4 and IL-13 and control profibrogenic macrophage program." in: Nature communications, Vol. 6, pp. 8523, (2015) (PubMed).

    García, Cidoncha, Bote, Hinchado, Ortega: "Altered profile of chemokines in fibromyalgia patients." in: Annals of clinical biochemistry, Vol. 51, Issue Pt 5, pp. 576-81, (2014) (PubMed).

    Wang, He, Tang, Zhang: "Chemokine expression in diverse nonimmediate drug hypersensitivity reactions: focus on thymus activation-regulated chemokine, cutaneous T-cell-attracting chemokine, and interleukin-10." in: Annals of allergy, asthma & immunology : official publication of the American College of Allergy, Asthma, & Immunology, Vol. 113, Issue 2, pp. 204-8, (2014) (PubMed).

    Shi, Song, Xiong, Cao, Liao, Ma, Zhang, Zeng, Liu, Wang, Cui, Huang, Liu: "Disease-specific T-helper cell polarizing function of lesional dendritic cells in different types of chronic rhinosinusitis with nasal polyps." in: American journal of respiratory and critical care medicine, Vol. 190, Issue 6, pp. 628-38, (2014) (PubMed).

    Jones, Vari, Keane, Crooks, Nourse, Seymour, Gottlieb, Ritchie, Gill, Gandhi: "Serum CD163 and TARC as disease response biomarkers in classical Hodgkin lymphoma." in: Clinical cancer research : an official journal of the American Association for Cancer Research, Vol. 19, Issue 3, pp. 731-42, (2013) (PubMed).

  • Target See all CCL17 ELISA Kits
    CCL17 (Chemokine (C-C Motif) Ligand 17 (CCL17))
    Alternative Name
    TARC (CCL17 Products)
    Synonyms
    A-152E5.3 ELISA Kit, ABCD-2 ELISA Kit, SCYA17 ELISA Kit, TARC ELISA Kit, CCL17 ELISA Kit, Abcd-2 ELISA Kit, Scya17 ELISA Kit, Scya17l ELISA Kit, Tarc ELISA Kit, abcd2 ELISA Kit, C-C motif chemokine ligand 17 ELISA Kit, chemokine (C-C motif) ligand 17 ELISA Kit, ATP-binding cassette sub-family D member 2 ELISA Kit, CCL17 ELISA Kit, Ccl17 ELISA Kit, LOC100537809 ELISA Kit
    Background
    The Human TARC (Thymus and Activation Regulated Chemokine) ELISA (Enzyme-Linked Immunosorbent Assay) kit is an in vitro enzyme-linked immunosorbent assay for the quantitative measurement of human TARC in serum, plasma, cell culture supernate and urine. This assay employs an antibody specific for human TARC coated on a 96-well plate. Standards and samples are pipetted into the wells and TARC present in a sample is bound to the wells by the immobilized antibody. The wells are washed and biotinylated anti-human TARC antibody is added. After washing away unbound biotinylated antibody, HRP-conjugated streptavidin is pipetted to the wells. The wells are again washed, a TMB substrate solution is added to the wells and color develops in proportion to the amount of TARC bound. The Stop Solution changes the color from blue to yellow, and the intensity of the color is measured at 450 nm. Reproducibility: Intra-Assay: CV<10% Inter-Assay: CV<12%.
    Gene ID
    6361
    UniProt
    Q92583
You are here:
Support