Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

GNRH1 ELISA Kit

GNRH1 Reactivity: Chicken Colorimetric Competition ELISA 31.25 pg/mL - 2000 pg/mL Plasma, Serum, Tissue Homogenate
Catalog No. ABIN5520316
  • Target See all GNRH1 ELISA Kits
    GNRH1 (Gonadotropin-Releasing Hormone 1 (Luteinizing-Releasing Hormone) (GNRH1))
    Reactivity
    • 3
    • 3
    • 3
    • 3
    • 2
    • 2
    • 2
    • 2
    • 2
    • 2
    • 2
    • 2
    • 1
    • 1
    Chicken
    Detection Method
    Colorimetric
    Method Type
    Competition ELISA
    Detection Range
    31.25 pg/mL - 2000 pg/mL
    Minimum Detection Limit
    31.25 pg/mL
    Application
    ELISA
    Purpose
    For quantitative detection of LHRH/GnRH in serum, plasma, tissue homogenates.
    Sample Type
    Plasma, Serum, Tissue Homogenate
    Analytical Method
    Quantitative
    Specificity
    This assay has high sensitivity and excellent specificity for detection of LHRH/GnRH. No significant cross- reactivity or interference between LHRH/GnRH and analogues was observed. Note: Limited by current skills and knowledge, it is difficult for us to complete the cross-reactivity detection between LHRH/GnRH and all the analogues, therefore, cross reaction may still exist.
    Sensitivity
    18.75 pg/mL
    Components
    • Pre-coated, ready to use 96-well strip plate
    • Plate sealer for 96 wells
    • Standard
    • Standard Dilution Buffer
    • Biotin-labeled Antibody (Concentrated)
    • Antibody Dilution Buffer
    • HRP-Streptavidin Conjugate (SABC)
    • SABC Dilution Buffer
    • TMB Substrate
    • Stop Solution
    • Wash Buffer (25 x concentrate)
    • Instruction manual
    Featured
    Discover our best selling GNRH1 ELISA Kit
    Top Product
    Discover our top product GNRH1 ELISA Kit
  • Application Notes
    Optimal working dilution should be determined by the investigator.
    Sample Volume
    50 μL
    Plate
    Pre-coated
    Protocol
    1. Wash plate 2 times before adding Standard, Sample and Control (blank) wells!
    2. Add 50 µL Standard or Sample into each well. Immediately add 50 µL Biotin-labeled Antibody into each well, gently tap the plate to ensure thorough mixing then incubate for 45 minutes at 37 °C.
    3. Aspirate and wash plates 3 times.
    4. Add 100 µL SABC Working Solution into each well and incubate for 30 minutes at 37 °C.
    5. Aspirate and wash plates 5 times.
    6. Add 90 µL TMB Substrate Solution. Incubate 10-20 minutes at 37 °C.
    7. Add 50 µL Stop Solution. Read at 450nm immediately and calculation.
    Reagent Preparation
    • Bring all reagents and samples to room temperature for 20 minutes before use.
    • Wash Buffer: If crystals have formed in the concentrate, you can warm it with 40 °C water bath Concentrated Wash Buffer into 750 mL Wash Buffer with deionized or distilled water. Put unused solution back at 2-8 °C.
    • Standards:
      1. Add 1 mL Sample Dilution Buffer into one Standard tube (labeled as zero tube), keep the tube at room temperature for 10 minutes and mix them thoroughly. Note: If the standard tube concentration higher than the range of the kit,please dilute it and labeled as zero tube.
      2. Label 7 EP tubes with 1/2, 1/4, 1/8, 1/16, 1/32, 1/64 and blank respectively. Add 0.3 mLof the Sample Dilution Buffer into each tube. Add 0.3 mLof the above Standard solution (from zero tube) into 1st tube and mix them thoroughly. Transfer 0.3 mL from 1st tube to 2nd tube and mix them thoroughly. Transfer 0.3 mL from 2nd tube to 3rd tube and mix them thoroughly, and so on. Sample Dilution Buffer was used for the blank control. Note: It is best to use Standard Solutions within 2 hours.
    • Preparation of Biotin-labeled Antibody Working Solution:
      Prepare it within 1 hour before experiment.
      1. Calculate required total volume of the working solution: 0.05ml/well x quantity of wells. (Allow 0.1-0.2 mLmore than the total volume.)
      2. Dilute the Biotin-detection antibody with Antibody Dilution Buffer at 1:100 and mix them thoroughly. (i.e. Add 1 µL Biotin-labeled antibody into 99 µL Antibody Dilution Buffer.)
    • Preparation of HRP-Streptavidin Conjugate (SABC) Working Solution:
      Prepare it within 30 minutes before experiment.
      1. Calculate required total volume of the working solution: 0.1ml/well x quantity of wells. (Allow 0.1-0.2 mLmore than the total volume.)
      2. Dilute the SABC with SABC Dilution Buffer at 1:100 and mix them thoroughly. (i.e. Add 1 µL of SABC into 99 µL of SABC Dilution Buffer.)
    Restrictions
    For Research Use only
  • Storage
    4 °C
    Storage Comment
    1. For unopened kit: All the reagents should be kept according to the labels on vials. The Reference Standard and the 96-well stripe plate should be stored at -20 °C upon receipt while the other reagents should be stored at 4 °C.
    2. For used kit: When the kit is used, the remaining reagents need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and zip-seal the foil pouch.
    Expiry Date
    6 months
  • Target See all GNRH1 ELISA Kits
    GNRH1 (Gonadotropin-Releasing Hormone 1 (Luteinizing-Releasing Hormone) (GNRH1))
    Alternative Name
    LHRH/GnRH (GNRH1 Products)
    Synonyms
    GNRH ELISA Kit, GRH ELISA Kit, HH12 ELISA Kit, LHRH ELISA Kit, LNRH ELISA Kit, Gnrh ELISA Kit, Gnrh2 ELISA Kit, Lhrh1 ELISA Kit, Lnrh ELISA Kit, hpg ELISA Kit, Gnrha ELISA Kit, Lhrh ELISA Kit, Rgnrhg1 ELISA Kit, CGNRH-I ELISA Kit, GnRH-I ELISA Kit, GnRH ELISA Kit, gnrh ELISA Kit, gnrh1 ELISA Kit, grh ELISA Kit, lhrh ELISA Kit, lnrh ELISA Kit, mdGnRH ELISA Kit, mdgnrh ELISA Kit, mfGnRH ELISA Kit, gonadotropin releasing hormone 1 ELISA Kit, gonadotropin releasing hormone 1 L homeolog ELISA Kit, prepro-gonadotropin-releasing hormone ELISA Kit, GNRH1 ELISA Kit, Gnrh1 ELISA Kit, gnrh1.L ELISA Kit, gnrh1 ELISA Kit
    UniProt
    P37042
    Pathways
    Hormone Activity
You are here: