Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

Transferrin ELISA Kit

TF Reactivity: Rat Colorimetric Sandwich ELISA Cell Culture Supernatant
Catalog No. ABIN612780
  • Target See all Transferrin (TF) ELISA Kits
    Transferrin (TF)
    Reactivity
    • 7
    • 5
    • 4
    • 3
    • 3
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Rat
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Minimum Detection Limit
    0.3 ng/mL
    Application
    ELISA
    Purpose
    The AssayMax Rat Transferrin ELISA kit is designed for detection of rat transferrin in urine and cell culture supernatants
    Brand
    AssayMax
    Sample Type
    Cell Culture Supernatant
    Analytical Method
    Quantitative
    Specificity
    10% FBS in culture media will not affect the assay.
    Components
    Rat Transferrin Microplate: A 96 well polystyrene microplate (12 strips of 8 wells) coated with a polyclonal antibody against Rat Transferrin. Sealing Tapes: Each kit contains 3 pre-cut, pressure-sensitive sealing tapes that can be cut to fit the format of the individual assay. Rat Transferrin Standard: Rat Transferrin in a buffered protein base (80 ng, lyophilized). 1 Biotinylated Transferrin Antibody (100x): A 100-fold concentrated biotinylated polyclonal antibody against transferrin (80µl). MIx Diluent Concentrate (10x): A 10-fold concentrated buffered protein base (30 ml). Wash Buffer Concentrate (20x): A 20-fold concentrated buffered surfactant (30 ml, 2 bottles). Streptavidin-Peroxidase Conjugate (SP Conjugate): A 100-fold concentrate (80µl). Chromogen Substrate: A ready-to-use stabilized peroxidase chromogen substrate tetramethylbenzidine (8 ml). Stop Solution: A 0.5 N hydrochloric acid to stop the chromogen substrate reaction (12 ml).
    Material not included
    Microplate reader capable of measuring absorbance at 450 nm. Pipettes (1-20 µL, 20-200 µL, 200-1000µLand multiple channel). Deionized or distilled reagent grade water.
    Featured
    Discover our best selling TF ELISA Kit
    Top Product
    Discover our top product TF ELISA Kit
  • Sample Volume
    50 μL
    Assay Time
    < 4 h
    Plate
    Pre-coated
    Protocol
    This assay employs a quantitative sandwich enzyme immunoassay technique that measures transferrin in less than 4 hours. A polyclonal antibody specific for transferrin has been pre-coated onto a microplate. Transferrin in standards and samples is sandwiched by the immobilized antibody and biotinylated polyclonal antibody specific for transferrin, which is recognized by a streptavidin-peroxidase conjugate. All unbound material is then washed away and a peroxidase enzyme substrate is added. The color development is stopped and the intensity of the color is measured.
    Reagent Preparation

    Freshly dilute all reagents and bring all reagents to room temperature before use. If crystals have formed in the concentrate, mix gently until the crystals have completely dissolved. MIx Diluent Concentrate (10x): Dilute the MIx Diluent 1:10 with reagent grade water. Store for up to 1 month at 2-8°C. Standard Curve: Reconstitute the 80 ng of Transferrin Standard with 4 ml of MIx Diluent to generate a standard solution of 20 ng/ml. Allow the standard to sit for 10 minutes with gentle agitation prior to making dilutions. Prepare duplicate or triplicate standard points by serially diluting the standard solution (20 ng/ml) 1:2 with MIx Diluent to produce 10, 5, 2.5, 1.25, 0.625, and 0.313 ng/ml solutions. MIx Diluent serves as the zero standard (0 ng/ml). Any remaining solution should be frozen at -20°C. Standard Point Dilution [R. Transferrin] (ng/ml) P1 Standard (20 ng/ml) 20.000 P2 1 part P1 + 1 part MIx Diluent 10.000 P3 1 part P2 + 1 part MIx Diluent 5.000 P4 1 part P3 + 1 part MIx Diluent 2.500 P5 1 part P4 + 1 part MIx Diluent 1.250 P6 1 part P5 + 1 part MIx Diluent 0.625 P7 1 part P6 + 1 part MIx Diluent 0.313 P8 MIx Diluent 0.000 Biotinylated Transferrin Antibody (100x): Spin down the biotinylated antibody briefly and dilute the desired amount of the antibody 1:100 with MIx Diluent. Any remaining solution should be frozen at -20°C. Wash Buffer Concentrate (20x): Dilute the Wash Buffer Concentrate 1:20 with reagent grade water. SP Conjugate (100x): Spin down the SP Conjugate briefly and dilute the desired amount of the conjugate 1:100 with MIx Diluent. Any remaining solution should be frozen at -20°C.

    Sample Collection
    Cell Culture Supernatants: Centrifuge cell culture media at 2000 x g for 10 minutes to remove debris. Collect supernatants and assay. Store samples at -20°C or below. Avoid repeated freeze-thaw cycles. Urine: Collect urine using sample pot. Centrifuge samples at 800 x g for 10 minutes and assay. Urine dilution is suggested at 1:200 in MIx Diluent, however, the user should determine the optimal dilution factor. The undiluted samples can be stored at -20°C or below for up to 3 months. Avoid repeated freeze-thaw cycles.
    Assay Procedure

    Prepare all reagents, working standards and samples as instructed. Bring all reagents to room temperature before use. The assay is performed at room temperature (20 - 30 °C). Remove excess microplate strips from the plate frame and return them immediately to the foil pouch with desiccant inside. Reseal the pouch securely to minimize exposure to water vapor and store in a vacuum desiccator. Add 50 µL of standard or sample per well. Cover wells with a sealing tape and incubate for two hours. Start the timer after the last sample addition. Wash five times with 200 µL of Wash Buffer manually. Invert the plate each time and decant the contents, hit it 4-5 times on absorbent paper towel to completely remove the liquid. If using a machine wash six times with 300 µL of Wash Buffer and then invert the plate, decant the contents, hit it 4-5 times on absorbent paper towel to completely remove the liquid. Add 50 µL of Biotinylated Transferrin Antibody to each well and incubate for one hour. Wash the microplate as described above. Add 50 µL of Streptavidin-Peroxidase Conjugate per well and incubate for 30 minutes. Turn on the microplate reader and set up the program in advance. Wash the microplate as described above. Add 50 µL of Chromogen Substrate per well and incubate for about 10 minutes or till the optimal blue color density develops. Gently tap plate to ensure thorough mixing and break the bubbles in the well with pipette tip. Add 50 µL of Stop Solution to each well. The color will change from blue to yellow. Read the absorbance on a microplate reader at a wavelength of 450 nm immediately. If wavelength correction is available, subtract readings at 570 nm from those at 450 nm to correct optical imperfections. Otherwise, read the plate at 450 nm only. Please note that some unstable black particles may be generated at high concentration points after stopping the reaction for about 10 minutes, which will reduce the readings. 3

    Calculation of Results

    Calculate the mean value of the triplicate readings for each standard and sample. To generate a Standard Curve, plot the graph using the standard concentrations on the x-axis and the corresponding mean 450 nm absorbance on the y-axis and draw a best fit curve through the points on the graph. Plotting the 4 parameters or linear graph may linearize the data and the best-fit line can be determined by regression analysis of the linear portion of the curve. Determine the unknown sample concentration from the Standard Curve and multiply the value by the dilution factor. Standard Curve The curve is provided for illustration only. A standard curve should be generated each time the assay is performed.

    Assay Precision
    Intra-assay and inter-assay coefficients of variation were 4.5% and 7.1% respectively.
    Restrictions
    For Research Use only
  • Handling Advice
    Prepare all reagents (working diluent buffer, wash buffer, standards, biotinylated- antibody, and SP conjugate) as instructed, prior to running the assay. Prepare all samples prior to running the assay. The dilution factors for the samples are suggested in this protocol. However, the user should determine the optimal dilution factor. Spin down the SP conjugate vial and the biotinylated-antibody vial before opening and using contents. The kit should not be used beyond the expiration date.
    Storage
    4 °C/-20 °C
    Storage Comment
    Store components of the kit at 2-8°C or -20°C upon arrival up to the expiration date. Store SP Conjugate and Biotinylated Antibody at -20°C Store Microplate, Diluent Concentrate (10x), Wash Buffer, Stop Solution, and Chromogen Substrate at 2-8°C Opened unused microplate wells may be returned to the foil pouch with the desiccant packs. Reseal along zip-seal. May be stored for up to 1 month in a vacuum desiccator. Diluent (1x) may be stored for up to 1 month at 2-8°C. Store Standard at 2-8°C before reconstituting with Diluent and at -20°C after reconstituting with Diluent.
  • Target See all Transferrin (TF) ELISA Kits
    Transferrin (TF)
    Alternative Name
    Transferrin (TF Products)
    Synonyms
    ltf ELISA Kit, pro1557 ELISA Kit, pro2086 ELISA Kit, mgc107777 ELISA Kit, LOC692564 ELISA Kit, TF ELISA Kit, LOC100144362 ELISA Kit, tf ELISA Kit, LTF ELISA Kit, TFEW ELISA Kit, conalbumin ELISA Kit, tf-b ELISA Kit, MGC64306 ELISA Kit, PRO1557 ELISA Kit, PRO2086 ELISA Kit, TFQTL1 ELISA Kit, AI266983 ELISA Kit, Cd176 ELISA Kit, HP ELISA Kit, Tf ELISA Kit, Tfn ELISA Kit, hpx ELISA Kit, Trf ELISA Kit, cb285 ELISA Kit, gavi ELISA Kit, id:ibd3238 ELISA Kit, id:ibd3525 ELISA Kit, sb:cb285 ELISA Kit, wu:fb57g06 ELISA Kit, wu:fb62h02 ELISA Kit, wu:fb63h10 ELISA Kit, wu:fb64h10 ELISA Kit, zgc:112154 ELISA Kit, 143958_at ELISA Kit, CG6186 ELISA Kit, Dmel\\CG6186 ELISA Kit, TSF1 ELISA Kit, anon-EST:Posey265 ELISA Kit, tsf1 ELISA Kit, Pro-TRH ELISA Kit, IL-5 ELISA Kit, STF I ELISA Kit, TRF1 ELISA Kit, sTF1 ELISA Kit, sTf ELISA Kit, tf1 ELISA Kit, transferrin ELISA Kit, serotransferrin ELISA Kit, transferrin (ovotransferrin) ELISA Kit, transferrin L homeolog ELISA Kit, melanotransferrin ELISA Kit, transferrin-a ELISA Kit, Transferrin 1 ELISA Kit, thyrotropin releasing hormone ELISA Kit, interleukin 5 ELISA Kit, TF ELISA Kit, LOC477072 ELISA Kit, tf ELISA Kit, Tf ELISA Kit, LOC100144362 ELISA Kit, tf.L ELISA Kit, LOC5575625 ELISA Kit, Trf ELISA Kit, tfa ELISA Kit, Tsf1 ELISA Kit, TRH ELISA Kit, IL5 ELISA Kit, LOC101085148 ELISA Kit, LOC100726872 ELISA Kit, trf ELISA Kit
    Background
    Transferrin is a plasma protein that transports iron through the blood to the liver, spleen and bone marrow. Low transferrin level in plasma could associate with anemia , and chronic liver disease. On the other hand, high plasma transferrin level could indicate iron deficiency anemia.
    Pathways
    Transition Metal Ion Homeostasis
You are here:
Support