Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

IL12A ELISA Kit

IL12A Reactivity: Mouse Colorimetric Sandwich ELISA 15.63 pg/mL - 1000 pg/mL Cell Culture Supernatant, Plasma, Serum
Catalog No. ABIN6963011
  • Target See all IL12A ELISA Kits
    IL12A (Interleukin 12 alpha (IL12A))
    Reactivity
    • 3
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Mouse
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Detection Range
    15.63 pg/mL - 1000 pg/mL
    Minimum Detection Limit
    15.63 pg/mL
    Application
    ELISA
    Purpose
    The kit is a sandwich enzyme immunoassay technique for the in vitro quantitative measurement in various sample types.
    Sample Type
    Cell Culture Supernatant, Plasma, Serum
    Analytical Method
    Quantitative
    Specificity
    This kit recognizes Mouse IL-12 p35 in samples. No Significant cross-reactivity or interference between Mouse IL-12 p35 and analogues was observed.
    Sensitivity
    9.38 pg/mL
    Components
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Reference Standard & Sample Diluent
    • Biotinylated Detection Antibody (100 x concentrate)
    • HRP Conjugate (100 x concentrate)
    • Biotinylated Detection Antibody Diluent
    • HRP Conjugate Diluent
    • Substrate Reagent
    • Stop Solution
    • Wash Buffer (25 x concentrate)
    • Instruction manual
    Top Product
    Discover our top product IL12A ELISA Kit
  • Sample Volume
    100 μL
    Assay Time
    3.5 h
    Plate
    Pre-coated
    Protocol
    1. Add 100 µL standard or sample to each well. Incubate for 90 min at 37 °C.
    2. Remove the liquid. Add 100 µL Biotinylated Detection Antibody. Incubate for 1 hour at 37 °C.
    3. Aspirate and wash 3 times.
    4. Add 100 µL HRP Conjugate. Incubate for 30 min at 37 °C.
    5. Aspirate and wash 5 times.
    6. Add 90 µL Substrate Reagent. Incubate for 15 min at 37 °C.
    7. Add 50 µL Stop Solution. Read at 450 nm immediately.
    8. Calculation of results.
    Reagent Preparation
    1. Bring all reagents to room temperature (18~25 °C) before use. Follow the Microplate reader manual for set-up and preheat it for 15 min before OD measurement.
    2. Wash Buffer: Dilute 30 mL of Concentrated Wash Buffer with 720 mL of deionized or distilled water to prepare 750 mL of Wash Buffer.Note: if crystals have formed in the concentrate, warm it in a 40 °C water bath and mix it gently until the crystals have completely dissolved
    3. Standard working solution: Centrifuge the standard at 10,000xg for 1 min. Add 1.0 mL of Reference Standard &Sample Diluent, let it stand for 10 min and invert it gently several times. After it dissolves fully, mix it thoroughly with a pipette. This reconstitution produces a working solution of 1000 pg/mL. Then make serial dilutions as needed. The recommended dilution gradient is as follows: 1000, 500, 250, 125, 62.5, 31.25, 15.63, 0 pg/mL. Dilution method: Take 7 EP tubes, add 500 μLof Reference Standard & Sample Diluent to each tube. Pipette 500 μLof the 1000 pg/mL working solution to the first tube and mix up to produce a 500 pg/mL working solution. Pipette 500 μLof the solution from the former tube into the latter one according to these steps. The illustration below is for reference. Note: the last tube is regarded as a blank. Don't pipette solution into it from the former tube.
    4. Biotinylated Detection Antibody working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Centrifuge the stock tube before use, dilute the 100x Concentrated Biotinylated Detection Antibody to 1xworking solution with Biotinylated Detection Antibody Diluent.
    5. Concentrated HRP Conjugate working solution: Calculate the required amount before the experiment (100 μL/well). In preparation, slightly more than calculated should be prepared. Dilute the 100x Concentrated HRP Conjugate to 1x working solution with Concentrated HRP Conjugate Diluent.
    Sample Preparation
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturationmay occur in these samples, leading to false results. Samples should therefore be stored for a short periodat 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thawcycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged toremove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determinecompatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibilityof causing a deviation due to the introduced chemical substance.The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in therange of the standard curve, the optimal sample dilution for the particular experiment has to be determined.Samples should then be diluted with PBS (pH =7.0-7.2).
    Assay Precision
    Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level Mouse IL-12 p35 were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level Mouse IL-12 p35 were tested on 3 different plates, 20 replicates in each plate.
    Both intra-CV and inter-CV are < 10 %.
    Restrictions
    For Research Use only
  • Storage
    4 °C,-20 °C
    Storage Comment
    1. For unopened kit: All reagents should be stored according to the labels on the vials, so they are stable up to 6 months after receipt of the kit. The Reference Standard, Biotinylated Detection Antibody, HRP Conjugate and the 96-well stripe plate should be stored at -20 °C upon receipt while the other reagents should be stored at 4 °C.
    2. For used kit: When the kit is used, the remaining reagents need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and zip-seal the foil pouch.
    Expiry Date
    6 months
  • Target See all IL12A ELISA Kits
    IL12A (Interleukin 12 alpha (IL12A))
    Alternative Name
    Interleukin 12 p35 (IL12A Products)
    Synonyms
    CLMF ELISA Kit, IL-12A ELISA Kit, NFSK ELISA Kit, NKSF1 ELISA Kit, P35 ELISA Kit, CLMF2 ELISA Kit, IL-12B ELISA Kit, NKSF ELISA Kit, NKSF2 ELISA Kit, p40 ELISA Kit, Il-12b ELISA Kit, Il12p40 ELISA Kit, Il-12p40 ELISA Kit, Il12 ELISA Kit, IL-12p40 ELISA Kit, IL-12p35 ELISA Kit, Il-12a ELISA Kit, Ll12a ELISA Kit, p35 ELISA Kit, IL12p35 ELISA Kit, Il-12 p35 ELISA Kit, IL12p40 ELISA Kit, IL12 ELISA Kit, IL-12 p35 ELISA Kit, interleukin 12A ELISA Kit, interleukin 12B ELISA Kit, interleukin 12b ELISA Kit, interleukin 12a ELISA Kit, Interleukin 12 level ELISA Kit, IL12A ELISA Kit, IL12B ELISA Kit, Il12b ELISA Kit, Il12a ELISA Kit, IL12 ELISA Kit, il12a ELISA Kit
    Background
    IL12A, CLMF1, NKSF1, p35, IL-12/IL-35 p35
    Pathways
    JAK-STAT Signaling, Regulation of Leukocyte Mediated Immunity, Positive Regulation of Immune Effector Process, Activated T Cell Proliferation