IL-6 ELISA Kit

IL6 Reactivity: Pig Colorimetric Sandwich ELISA 31.2-2000 pg/mL
Catalog No. ABIN955914
  • Target See all IL-6 (IL6) ELISA Kits
    IL-6 (IL6) (Interleukin 6 (IL6))
    Reactivity
    • 14
    • 10
    • 9
    • 6
    • 6
    • 4
    • 4
    • 3
    • 3
    • 3
    • 2
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    • 1
    • 1
    • 1
    Pig
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Detection Range
    31.2-2000 pg/mL
    Minimum Detection Limit
    31.2 pg/mL
    Application
    ELISA
    Purpose
    The kit is a sandwich enzyme immunoassay for the in vitro quantitative measurement of pig IL6 in serum, plasma, tissue homogenates, cell culture supernates and other biological fluids.
    Analytical Method
    Quantitative
    Specificity
    This assay has high sensitivity and excellent specificity for detection of pig IL6. No significant cross- reactivity or interference between pig IL6 and analogues was observed. Note: Limited by current skills and knowledge, it is impossible for us - complete the cross-reactivity detection between pig IL6 and all the analogues, therefore, cross reaction may still exist.
    Sensitivity
    The minimum detectable dose of pig IL6 is typically less than 12.3 pg/mL.
    The Sensitivity of this assay, or Lower Limit of Detection (LLD) was defined as the lowest protein concentration that could be differentiated from zero. It was determined the mean O.D. Value of 20 replicates of the zero calibrator plus three standard deviations.
    Characteristics
    The microtiter plate provided in this kit has been pre-coated with a monoclonal antibody specific to IL6. Calibrators or samples are then added to the appropriate microtiter plate wells with a biotin-conjugated polyclonal antibody preparation specific for IL6. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. Then a TMB substrate solution is added to each well. Only those wells that contain IL6, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of a sulfuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450 nm +/- 10 nm. The concentration of IL6 in the samples is then determined by comparing the O.D. of the samples to the calibration curve.
    Components
    Pre-coated, ready to use 96-well strip plate (1x)
    Calibrator (lyophilized) (2x)
    Calibrator Diluent (1 x 20 mL)
    Detection Reagent A (1 x 120 µL)
    Detection Reagent B (1 x 120 µL)
    Assay Diluent A (2X concentrate) (1 x 6 mL)
    Assay Diluent B (2X concentrate) (1 x 6 mL)
    TMB Substrate (1 x 9 mL)
    Stop Solution (1 x 6 mL)
    Wash Buffer (30X concentrate) (1 x 20 mL)
    Plate sealer for 96 wells (4x).
    Material not included
    1. Microplate reader with 450 +/- 10 nm filter.
    2. Precision single or multi-channel pipettes and disposable tips.
    3. Eppendorf Tubes for diluting samples.
    4. De-ionized or distilled water.
    5. Absorbent paper for blotting the microtiter plate.
    6. Container for Wash Solution.
    Featured
    Discover our best selling IL6 ELISA Kit
    Top Product
    Discover our top product IL6 ELISA Kit
  • Comment

    The calibration curve concentrations used for the ELISA's were 2, 000 pg/mL, 1000 pg/mL, 500 pg/mL, 250 pg/mL, 125 pg/mL, 62.5 pg/mL, 31.2 pg/mL.

    Plate
    Pre-coated
    Calculation of Results

    Average the duplicate readings for each calibrator, control, and samples and subtract the average zero calibrator optical density. Create a calibration curve on log-log graph paper, with IL6 concentration on the y-axis and absorbance on the x-axis. Draw the best fit straight line through the calibrator points and it can be determined by regression analysis. Using some plot software is also recommended. If samples have been diluted, the concentration read from the calibration curve must be multiplied by the dilution factor.

    Restrictions
    For Research Use only
  • Storage
    -20 °C
    Storage Comment
    All the reagents should be kept according to the labels on vials. The Calibrator, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20° C upon being received. The unused strips should be kept in a sealed bag with the desiccant provided to minimize exposure to damp air. Opened test kits will remain stable until the expiration date shown, provided it is stored as prescribed above.
    Expiry Date
    The expiry date is stated on the label.
  • Target See all IL-6 (IL6) ELISA Kits
    IL-6 (IL6) (Interleukin 6 (IL6))
    Abstract
    IL6 Products
    Synonyms
    BSF2 ELISA Kit, HGF ELISA Kit, HSF ELISA Kit, IFNB2 ELISA Kit, IL-6 ELISA Kit, Il-6 ELISA Kit, ILg6 ELISA Kit, Ifnb2 ELISA Kit, il6 ELISA Kit, CHIL-6 ELISA Kit, interleukin 6 ELISA Kit, interleukin-6 ELISA Kit, IL6 ELISA Kit, Il6 ELISA Kit, il-6 ELISA Kit, IL-6 ELISA Kit
    Pathways
    TLR Signaling, Hormone Transport, Negative Regulation of Hormone Secretion, Myometrial Relaxation and Contraction, Positive Regulation of Immune Effector Process, Production of Molecular Mediator of Immune Response, Regulation of Carbohydrate Metabolic Process, Autophagy, Cell RedoxHomeostasis, Cancer Immune Checkpoints, Inflammasome