Titin ELISA Kit

This Human Titin ELISA Kit is a Colorimetric ELISA Kit designed to quantify Human Titin.
Catalog No. ABIN454416

Quick Overview for Titin ELISA Kit (ABIN454416)

Target

See all Titin (TTN) ELISA Kits
Titin (TTN)

Reactivity

  • 3
  • 1
  • 1
Human

Detection Method

Colorimetric

Method Type

Sandwich ELISA

Detection Range

0.312 ng/mL - 20 ng/mL

Application

ELISA

Sample Type

Cell Culture Supernatant, Plasma, Serum, Tissue Homogenate, Urine
  • Minimum Detection Limit

    0.312 ng/mL

    Purpose

    This immunoassay kit allows for the in vitro quantitative determination of human Titin , concentrations in serum, Plasma, Urine, tissue homogenates and Cell culture supernates.

    Analytical Method

    Quantitative

    Specificity

    This assay recognizes recombinant and natural human Anti-titin Ab.

    Cross-Reactivity (Details)

    No significant cross-reactivity or interference was observed.

    Sensitivity

    0.138 ng/mL

    Characteristics

    Homo sapiens,Human,Titin,Connectin,Rhabdomyosarcoma antigen MU-RMS-40.14,TTN,2.7.11.1

    Components

    Reagent (Quantity): Assay plate (1), Standard (2), Sample Diluent (1x20ml), Assay Diluent A (1x10ml), Assay DiluentB 1 x 10ml Detection Reagent A (1x120µl), Detection Reagent B (1x120µl), Wash Buffer(25 x concentrate) (1x30ml), Substrate (1x10ml), Stop Solution (1x10ml)
  • Sample Volume

    100 μL

    Plate

    Pre-coated

    Reagent Preparation

    Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Dilute 30 mL of Wash Buffer Concentrate into deionized or distilled water to prepare 750 mL of Wash Buffer. Standard - Reconstitute the Standard with 1.0mL of Sample Diluent. This reconstitution produces a stock solution of 20ng/mL. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making serial dilutions (Making serial dilution in the wells directly is not permitted). The undiluted standard serves as the high standard (20ng/mL) The Sample Diluent serves as the zero standard (0ng/mL). Detection Reagent A and B - Dilute to the working concentration using Assay Diluent A and B (1:100), respectively.

    Sample Collection

    Serum - Use a serum separator tube (SST) and allow samples to clot for 30 minutes before centrifugation for 15 minutes at approximately 1000 x g. Remove serum and assay immediately or aliquot and store samples at -20 °C or -80 °C. Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 minutes at 1000 x g at 2 - 8 °C within 30 minutes of collection. Store samples at -20 °C or -80° 2 C. Avoid repeated freeze-thaw cycles. Other biological fluids - Remove particulates by centrifugation and assay immediately or aliquot and store samples at -20 °C or -80 °C. Avoid repeated freeze-thaw cycles. Note: Serum, plasma, and cell culture supernatant samples to be used within 7 days may be stored at 2-8C, otherwise samples must stored at -20 °C (≤ 3 months) or -80 °C (≤ 6 months) to avoid loss of bioactivity and contamination. Avoid freeze-thaw cycles. When performing the assay slowly bring samples to room temperature. It is recommended that all samples be assayed in duplicate.

    Calculation of Results

    Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the ANTI-TITIN AB concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.

    Assay Precision

    Intra-AssayCV: < 3.9%Inter-AssayCV: < 6.2%

    Restrictions

    For Research Use only
  • Handling Advice

    1. The kit should not be used beyond the expiration date on the kit label.
    2. Do not mix or substitute reagents with those from other lots or sources.
    3. If samples generate values higher than the highest standard, further dilute the samples with the Assay Diluent and repeat the assay. Any variation in standard diluent, operator, pipetting technique, washing technique,incubation time or temperature, and kit age can cause variation in binding.
    4. This assay is designed to eliminate interference by soluble receptors, ligands, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Immunoassay, the possibility of interference cannot be excluded.

    Storage

    4 °C/-20 °C

    Storage Comment

    The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 °C upon being received. The other reagents can be stored at 4 °C.
  • Target See all Titin (TTN) ELISA Kits

    Titin (TTN)

    Alternative Name

    TTN

    Target Type

    Antibody