IgA ELISA Kit

This Colorimetric ELISA kit is designed for the quantitative measurement of Human IgA.
Catalog No. ABIN3198549
$559.10
Plus shipping costs $50.00
96 tests
Shipping to: United States
Delivery in 3 to 5 Business Days

Quick Overview for IgA ELISA Kit (ABIN3198549)

Target

See all IgA ELISA Kits
IgA

Reactivity

  • 92
  • 22
  • 20
  • 13
  • 10
  • 10
  • 10
  • 5
  • 5
  • 3
  • 3
  • 3
  • 3
  • 2
  • 1
  • 1
Human

Detection Method

Colorimetric

Method Type

Sandwich ELISA

Detection Range

3.13 ng/mL - 200 ng/mL

Application

ELISA

Sample Type

Cell Culture Supernatant, Plasma, Serum
  • Minimum Detection Limit

    3.13 ng/mL

    Purpose

    This assay employs the quantitative sandwich enzyme immunoassay technique for quantitative detection.

    Analytical Method

    Quantitative

    Specificity

    IgA ELISA Kit, Human

    Components

    • Plate
    • Standard
    • Diluent

    Material not included

    Microplate reader capable of measuring absorbance at 450 nm, with correction wavelength set at 570 nm or 630 nm.
    Pipettes and pipette tips.
    50 μL to 300 μL adjustable multichannel micropipette with disposable tips.
    Multichannel micropipette reservoir.
    Beakers, flasks, cylinders necessary for preparation of reagents.
    Deionized or distilled water.
    Polypropylene test tubes for dilution.
  • Application Notes

    Optimal working dilution should be determined by the investigator.

    Sample Volume

    100 μL

    Assay Time

    3 - 4 h

    Plate

    Pre-coated

    Restrictions

    For Research Use only
  • Preservative

    Sodium azide

    Precaution of Use

    This product contains Sodium azide: a POISONOUS AND HAZARDOUS SUBSTANCE which should be handled by trained staff only.

    Handling Advice

    protect your eyes

    Storage

    4 °C

    Storage Comment

    Store kit reagents between 2 and 8 °C. Immediately after use remaining reagents should be returned to cold storage (2 to 8 °C).
    Expiry of the kit and reagents is stated on labels. Expiration date of the kit components can only be guaranteed if the components are stored properly, and if, in case of repeated use of one component, this reagent is not contaminated by the first handling.
    Unopened Kit: Store at 2 - 8 °C (See expiration date on the label).
    Opened/Reconstituted Reagents: Up to 1 month at 2 - 8 °C.
    Reconstituted Standard: Up to 1 month at ≤ -20 °C in a manual defrost freezer. Avoid repeated freeze-thaw cycles.
    Microplate Wells: Up to 1 month at 2 - 8 °C. Return unused strips to the foil pouch containing the desiccant pack, reseal along entire edge to maintain plate integrity. Provided this is within the expiration date of the kit
  • Target See all IgA ELISA Kits

    IgA

    Target Type

    Antibody

    Background

    Immunoglobulin A (IgA) is an antibody that plays a critical role in mucosal immunity. Its heavy chains are of the type α. IgA has two subclasses (IgA1 and IgA2) and can exist in a dimeric form called secretory IgA (sIgA). In its secretory form, IgA is the main immunoglobulin found in mucous secretions, it is also found in small amounts in blood. The secretory component of sIgA protects the immunoglobulin from being degraded by proteolytic enzymes, thus sIgA can survive in the harsh gastrointestinal tract environment and provide protection against microbes that multiply in body secretions. sIgA can also inhibit inflammatory effects of other immunoglobulins. IgA is a poor activator of the complement system, and opsonises only weakly.