SPTAN1 ELISA Kit

SPTAN1 Reactivity: Mouse Colorimetric Sandwich ELISA Plasma, Serum
Catalog No. ABIN454293
  • Target See all SPTAN1 ELISA Kits
    SPTAN1 (Spectrin alpha Chain, Brain (SPTAN1))
    Reactivity
    • 3
    • 1
    • 1
    • 1
    Mouse
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Application
    ELISA
    Purpose
    This immunoassay kit allows for the specific measurement of mouse Alpha-Fodrin IgG/IgA concentrations in serum and plasma.
    Sample Type
    Serum, Plasma
    Analytical Method
    Quantitative
    Specificity
    This assay recognizes recombinant and natural mouse Alpha-Fodrin IgG/IgA.
    Cross-Reactivity (Details)
    No significant cross-reactivity or interference was observed.
    Characteristics
    Mus musculus,Mouse,Spectrin alpha chain, brain,Alpha-II spectrin,Fodrin alpha chain,Spectrin, non-erythroid alpha chain,Sptan1,Spna2,Spta2
    Components
    Reagent (Quantity): Assay plate (1), Standard (2), Sample Diluent (1x20ml), 2 Assay Diluent A (1x10ml), Assay Diluent B (1x10ml), Detection Reagent A 1×120µl Detection Reagent B 1×120µl Wash Buffer(25 x concentrate) (1x30ml), Substrate (1x10ml), Stop Solution (1x10ml)
    Featured
    Discover our best selling SPTAN1 ELISA Kit
    Top Product
    Discover our top product SPTAN1 ELISA Kit
  • Sample Volume
    100 μL
    Plate
    Pre-coated
    Protocol
    This assay employs the quantitative sandwich enzyme immunoassay technique. A monoclonal antibody specific for Alpha-Fodrin IgG/IgA has been pre-coated onto a microplate. Standards and samples are pipetted into the wells and any Alpha-Fodrin IgG/IgA present is bound by the immobilized antibody. An enzyme-linked monoclonal antibody specific for Alpha-Fodrin IgG/IgA is added to the wells. Following a wash to remove any unbound antibody-enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Alpha-Fodrin IgG/IgA bound in the initial step. The color development is stopped and the intensity of the color is measured.
    Reagent Preparation

    Bring all reagents to room temperature before use. Wash Buffer - If crystals have formed in the concentrate, warm to room temperature and mix gently until the crystals have completely dissolved. Dilute 20 mL of Wash Buffer Concentrate into deionized or distilled water to prepare 500 mL of Wash Buffer. Standard - Reconstitute the Standard with 1.0 mL of Sample Diluent. This reconstitution produces a stock solution of 1000 U/mL. Allow the standard to sit for a minimum of 15 minutes with gentle agitation prior to making serial dilutions. The undiluted standard serves as the high standard (1000 U/mL). The Sample Diluent serves as the zero standard (0 U/mL). Detection Reagent A and B - Dilute to the working concentration specified on the vial label using Assay Diluent A and B (1:100), respectively.

    Sample Collection
    Serum - Use a serum separator tube (SST) and allow samples to clot for 30 minutes before centrifugation for 15 minutes at approximately 1000 x g. Remove serum and assay immediately or aliquot and store samples at -20 °C. Plasma - Collect plasma using EDTA or heparin as an anticoagulant. Centrifuge samples for 15 minutes at 1000 x g at 2 - 8 °C within 30 minutes of collection. Store samples at ≤ -20 °C. Avoid repeated freeze-thaw cycles. Note: Citrate plasma has not been validated for use in this assay.
    Assay Procedure

    Allow all reagents to reach room temperature. Arrange and label required number of strips. 3
    1. Prepare all reagents, working standards and samples as directed in the previous sections.
    2. Add 100 uL of Standard, Control, or sample per well. Cover with the adhesive strip. Incubate for 2 hours at 37 °C.
    3. Remove the liquid of each well, don’t wash.
    4. Add 100 uL of Detection Reagent A to each well. Incubate for 1 hour at 37°C. Detection Reagent A may appear cloudy. Warm to room temperature and mix gently until solution appears uniform.
    5. Aspirate each well and wash, repeating the process three times for a total of three washes. Wash by filling each well with Wash Buffer (350 uL) using a squirt bottle, multi-channel pipette, manifold dispenser or autowasher. Complete removal of liquid at each step is essential to good performance. After the last wash, remove any remaining Wash Buffer by aspirating or decanting. Invert the plate and blot it against clean paper towels.
    6. Add 100 uL of Detection Reagent B to each well. Cover with a new adhesive strip.Incubate for 1 hours at 37 °C.
    7. Repeat the aspiration/wash as in step
    5. 8. Add 90 uL of Substrate Solution to each well. Incubate for 30 minutes at room temperature. Protect from light.
    9. Add 50 uL of Stop Solution to each well. If color change does not appear uniform, gently tap the plate to ensure thorough mixing.
    10. Determine the optical density of each well within 30 minutes, using a microplate reader set to 450 nm.
    Important Note:
    1. The wash procedure is critical. Insufficient washing will result in poor precision and falsely elevated absorbance readings.
    2. It is recommended that no more than 32 wells be used for each assay run if manual pipetting is used since pipetting of all standards, specimens and controls should be completed within 5 minutes. A full plate of 96 wells may be used if automated pipetting is available.
    3. Duplication of all standards and specimens, although not required, is recommended.
    4. When mixing or reconstituting protein solutions, always avoid foaming.
    5. To avoid cross-contamination, change pipette tips between additions of each standard level, between sample additions, and between reagent additions. Also, use separate reservoirs for each reagent.
    6. To ensure accurate results, proper adhesion of plate sealers during incubation steps is necessary.

    Calculation of Results

    Average the duplicate readings for each standard, control, and sample and subtract the average zero standard optical density. Create a standard curve by reducing the data using computer software capable of generating a four parameter logistic (4-PL) curve-fit. As an alternative, 4 construct a standard curve by plotting the mean absorbance for each standard on the y-axis against the concentration on the x-axis and draw a best fit curve through the points on the graph. The data may be linearized by plotting the log of the ALPHA-FODRIN IGG/IGA concentrations versus the log of the O.D. and the best fit line can be determined by regression analysis. This procedure will produce an adequate but less precise fit of the data. If samples have been diluted, the concentration read from the standard curve must be multiplied by the dilution factor.

    Restrictions
    For Research Use only
  • Handling Advice
    1. The kit should not be used beyond the expiration date on the kit label.
    2. Do not mix or substitute reagents with those from other lots or sources.
    3. If samples generate values higher than the highest standard, further dilute the samples with the Assay Diluent and repeat the assay. Any variation in standard diluent, operator, pipetting technique, washing technique,incubation time or temperature, and kit age can cause variation in binding.
    4. This assay is designed to eliminate interference by soluble receptors, ligands, binding proteins, and other factors present in biological samples. Until all factors have been tested in the Immunoassay, the possibility of interference cannot be excluded.
    Storage
    4 °C/-20 °C
    Storage Comment
    The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20 °C upon being received. The other reagents can be stored at 4 °C.
  • Target See all SPTAN1 ELISA Kits
    SPTAN1 (Spectrin alpha Chain, Brain (SPTAN1))
    Alternative Name
    Sptan1 (SPTAN1 Products)
    Synonyms
    im:7157190 ELISA Kit, wu:fa20e05 ELISA Kit, wu:fb33g08 ELISA Kit, wu:fk32a10 ELISA Kit, zgc:112229 ELISA Kit, 2610027H02Rik ELISA Kit, Spna-2 ELISA Kit, Spna2 ELISA Kit, EIEE5 ELISA Kit, NEAS ELISA Kit, SPTA2 ELISA Kit, A2a ELISA Kit, IPF ELISA Kit, SPECA ELISA Kit, spectrin alpha, non-erythrocytic 1 ELISA Kit, spectrin, alpha, non-erythrocytic 1 ELISA Kit, SPTAN1 ELISA Kit, sptan1 ELISA Kit, Sptan1 ELISA Kit
    Background
    Alpha-Fodrin is an intracellular, actin-binding, organ-specific protein of the cytoskeleton. It is a dimer composed of an alpha- and a beta-subunit. The network of actin and fodrin situated below the plasma membrane of secretorial cells, is important for the alignment of secretorial vesicles to the plasma membrane during secretorial processes. During apoptosis the alpha-fodrin dimer is cleaved into a 120 kDa breakdown product, which is found abundantly in the salivary gland. This proteolysis of fodrin may be a consequence of protease activation during apoptosis. The cleavage product of 120 kDa alpha-fodrin was found to be an important autoantigen in the pathogenesis of organ-specific autoimmune response. Clinical studies have shown, that in patients with Sjögren Syndrome alpha-fodrin is involved in the stimulation of peripheral blood T-cells. These findings suggest, that an increase in protease activity and the stimulation of T-cells play an important role in the alpha-fodrin proteolysis during the development of primary Sjogren's Syndrome .
    Pathways
    Caspase Cascade in Apoptosis, Regulation of Actin Filament Polymerization
You are here:
Support