Mast Cell Tryptase ELISA Kit
Quick Overview for Mast Cell Tryptase ELISA Kit (ABIN5520414)
Target
See all Mast Cell Tryptase ELISA KitsReactivity
Detection Method
Method Type
Detection Range
Application
Sample Type
-
-
Minimum Detection Limit
- 0.156 ng/mL
-
Purpose
- Human TPS (Tryptase) ELISA Kit
-
Analytical Method
- Quantitative
-
Specificity
- Specifically binds with TPS , no obvious cross reaction with other analogues.
-
Sensitivity
- 0.094 ng/mL
-
Components
-
- ELISA Microplate(Dismountable)
- Lyophilized Standard
- HRP Antibody(Concentrated
- 100X)
- TMB Substrate
- Sample Dilution Buffer
- Stop Solution
- Plate Sealer
- Product Description
-
Material not included
-
- Microplate reader (wavelength: 450nm)
- 37 °C incubator (CO2 incubator for cell culture is not recommenced.)
- Automated plate washer or multi-channel pipette/5 mL pipettor
- Precision single (0.5-10μL
- 5-50μL
- 20-200μL
- 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
- Sterile tubes and Eppendorf tubes with disposable tips
- Absorbent paper and loading slot
- Deionized or distilled water
-
-
-
-
Application Notes
- Optimal working dilution should be determined by the investigator.
-
Comment
-
Sandwich ELISA, Double Antibody
-
Sample Volume
- 100 μL
-
Assay Time
- 4 h
-
Plate
- Pre-coated
-
Protocol
- Step 1: Wash the plate twice before adding the standard and sample. Step 2: Add 100 µL standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37 °C. Washing: Wash the plate three times and immerse for 1 min each time. Step 3: Add 100 µL HRP-antibody working solution, seal the plate and statically incubate for 30 minutes at 37 °C. Washing: Wash the plate five times and immerse for 1 min each time. Step 4: Add 90 µL TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37 °C. (Accurate TMB visualization control is required.) Step 5: Add 50 µL stop solution. Read at 450nm immediately and calculate.
-
Restrictions
- For Research Use only
-
-
-
Handling Advice
- Avoid direct light exposure to antibody solutions
-
Storage
- 4 °C
-
Storage Comment
- Store sealed Kit at 2-8° C. Do Not Cryopreserve. Store ELISA Microplate for 1 month at 2-8°C, Store for 12 month at -20°C. Put the leftover strips into a sealed foil bag with the desiccant
-
Expiry Date
- 12 months
-
-
- Mast Cell Tryptase
-
Alternative Name
- TPS
-
Background
- TPS, Tryptase