AMH CLIA Kit
Quick Overview for AMH CLIA Kit (ABIN5651851)
Target
See all AMH CLIA KitsReactivity
Detection Method
Method Type
Detection Range
Application
Sample Type
-
-
Minimum Detection Limit
- 195.3 pg/mL
-
Purpose
- Human Anti-Mullerian Hormone (AMH) CLIA Kit for detecting AMH in serum, plasma
-
Analytical Method
- Quantitative
-
Specificity
- This assay has high sensitivity and excellent specificity for detection of Anti-Mullerian Hormone (AMH). No significant cross-reactivity or interference between Anti-Mullerian Hormone (AMH) and analogues was observed.
-
Sensitivity
- 74.3 pg/mL
-
Components
-
- Pre-coated, ready to use 96-well strip plate
- Standar
- Detection Reagent A
- Detection Reagent B
- Substrate A
- Substrate B
- Assay Diluent A
- Assay Diluent B
- Wash Buffer (30x concentrate)
- Standard Diluent
- Plate Sealer
- Instruction Manual
-
Material not included
-
- Microplate Reader with 450 nm filter
- Single or multi-channel pipettes with high precision and disposable tips
- Microcentrifuge tubes
- Deionized or distilled watter
- Absorbent paper for blotting the microplate
- Container for wash solution
- 0.01 M (or 1x) Phosphate Buffered Saline (PBS), ph 7.0-7.2
-
-
-
-
Application Notes
- Optimal working dilution should be determined by the investigator.
-
Comment
-
This product is tested and assembled before shipping to ensure that it meets its quality specifications. All kits are tested to confirm that they fall within their defined Inter- and Intra- assay coefficient of variation.
-
Sample Volume
- 50 μL
-
Assay Time
- 2 h
-
Plate
- Pre-coated
-
Protocol
- The microplate provided in this kit has been pre-coated with a monoclonal antibody specific to Anti-Mullerian Hormone (AMH). A competitive inhibition reaction is launched between biotin labeled Anti-Mullerian Hormone (AMH) and unlabeled Anti-Mullerian Hormone (AMH) (Standards or samples) with the pre-coated antibody specific to Anti-Mullerian Hormone (AMH). After incubation the unbound conjugate is washed off. Next, avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. The amount of bound HRP conjugate is reverse proportional to the concentration of Anti-Mullerian Hormone (AMH) in the sample. Then the mixture of substrate A and B is added to generate glow light emission kinetics. Upon plate development, the intensity of the emitted light is reverse proportional to the Anti-Mullerian Hormone (AMH) level in the sample or standard.
-
Assay Precision
-
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Anti-Mullerian Hormone (AMH) were tested 20 times on one plate, respectively
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Anti-Mullerian Hormone (AMH) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12% -
Restrictions
- For Research Use only
-
-
-
Handling Advice
- Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.
-
Storage
- 4 °C,-20 °C
-
Storage Comment
- -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.
-
Expiry Date
- 12 months
-
-
- AMH (Anti-Mullerian Hormone (AMH))
-
Alternative Name
- Anti-Mullerian Hormone
-
Background
- Gene Aliases: MIF, MIH, MIS, Müllerian Inhibiting Factor, Müllerian Inhibiting Hormone, Müllerian Inhibiting Substance
-
Gene ID
- 268
-
UniProt
- P03971
-
Pathways
- Negative Regulation of Hormone Secretion
Target See all AMH CLIA Kits
-