Anti-Sperm Antibody (AsAb) ELISA Kit

AsAb Reactivity: Human Colorimetric Sandwich ELISA 3.12 ng/mL - 200 ng/mL Plasma, Serum
Catalog No. ABIN5651857
  • Target
    Anti-Sperm Antibody (AsAb)
    Reactivity
    Human
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Detection Range
    3.12 ng/mL - 200 ng/mL
    Minimum Detection Limit
    3.12 ng/mL
    Application
    ELISA
    Sample Type
    Plasma, Serum
    Analytical Method
    Quantitative
    Specificity
    This assay has high sensitivity and excellent specificity for detection of Anti-Sperm Antibody (AsAb). No significant cross-reactivity or interference between Anti-Sperm Antibody (AsAb) and analogues was observed.
    Sensitivity
    1.17 ng/mL
    Featured
    Discover our best selling AsAb ELISA Kit
  • Comment

    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5 % within the expiration date under appropriate storage condition. To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

    Assay Time
    2 - 3 h
    Plate
    Pre-coated
    Protocol
    This assay employs the competitive inhibition enzyme immunoassay technique. An antibody specific to Anti-Sperm Antibody (AsAb) has been pre-coated onto a microplate. A competitive inhibition reaction is launched between Horseradish Peroxidase (HRP) labeled Anti-Sperm Antibody (AsAb) and unlabeled Anti-Sperm Antibody (AsAb) (Standards or samples) with the pre-coated antibody specific to Anti-Sperm Antibody (AsAb). After incubation the unbound conjugate is washed off. The amount of bound HRP conjugate is reverse proportional to the concentration of Anti-Sperm Antibody (AsAb) in the sample. After addition of the substrate solution, the intensity of color developed is reverse proportional to the concentration of Anti-Sperm Antibody (AsAb) in the sample.
    Assay Precision
    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Anti-Sperm Antibody (AsAb) were tested 20 times on one plate, respectively
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Anti-Sperm Antibody (AsAb) were tested on 3 different plates, 8 replicates in each plate. CV(%) = SD/meanX100
    Intra-Assay: CV<10%
    Inter-Assay: CV<12%
    Restrictions
    For Research Use only
  • Handling Advice
    The Stop Solution is acidic. Do not allow to contact skin or eyes. Calibrators, controls and specimen samples should be assayed in duplicate. Once the procedure has been started, all steps should be completed without interruption.
    Storage
    4 °C,-20 °C
    Storage Comment
    -20°C. Bring all reagents to room temperature before beginning test. The kit may be stored at 4°C for immediate use within two days upon arrival. Reseal any unused strips with desiccant pack. Minimize freeze/thaw cycles.
    Expiry Date
    4-8 months
  • Target
    Anti-Sperm Antibody (AsAb)
    Alternative Name
    Anti-Sperm Antibody
    Target Type
    Antibody
    Background

    Gene Name: Anti-Sperm Antibody

    Gene Aliases: Anti-Spermatozoa Antibodies, Sperm Antibodies, Antisperm Antibodies