Quick Overview for Spermidine CLIA Kit (ABIN8056037)

Target

See all Spermidine products
Spermidine

Reactivity

Various Species

Detection Method

Chemiluminescent

Method Type

Competition ELISA

Detection Range

781.25 pg/mL - 200000 pg/mL

Application

ELISA

Sample Type

Cell Culture Supernatant, Cell Lysate, Plasma, Serum, Tissue Homogenate
  • Minimum Detection Limit

    781.25 pg/mL

    Purpose

    CLIA Kit for Spermidine (SMD)

    Analytical Method

    Quantitative

    Specificity

    This assay has high sensitivity and excellent specificity for detection of Spermidine (SMD).
    No significant cross-reactivity or interference between Spermidine (SMD) and analogues was observed.

    Sensitivity

    < 284.75 pg/mL

    Components

    • Pre-coated, 96-well strip plate
    • Plate sealer
    • Standard
    • Standard Diluent
    • Detection Reagent A
    • Assay Diluent A
    • Detection Reagent B
    • Assay Diluent B
    • Substrate A
    • Substrate B
    • Wash Buffer (30X)
    • Instruction manual

    Material not included

    • Luminometer capable of reading 96-well microplates with the following parameters: lag time 30.0secs; read time 1.0 sec/well
    • Single or multi-channel pipettes with high precision and disposable tips.
    • Microcentrifuge Tubes.
    • Deionized or distilled water.
    • Absorbent paper for blotting the microplate.
    • Container for Wash Solution.
    • 0.01mol/L (or 1×) Phosphate Buffered Saline(PBS), pH7.0-7.2
  • Sample Volume

    50 μL

    Assay Time

    2 h

    Plate

    Pre-coated

    Protocol

    1. Prepare all reagents, samples and standards;
    2. Add 50µL standard or sample to each well.
        And then add 50µL prepared Detection Reagent A immediately.
        Shake and mix. Incubate 1 hour at 37°C;
    3. Aspirate and wash 3 times;
    4. Add 100µL prepared Detection Reagent B. Incubate 30 minutes at 37°C;
    5. Aspirate and wash 5 times;
    6. Add 100µL Substrate Solution. Incubate 10 minutes at 37°C;
    7. Read RLU value immediately.

    Assay Precision

    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Spermidine (SMD) were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Spermidine (SMD) were tested on 3 different plates, 8 replicates in each plate.
    CV(%) = SD/meanX100
    Intra-Assay: CV<10%
    Inter-Assay: CV<12%

    Restrictions

    For Research Use only
  • Handling Advice

    The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
    To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.

    Storage

    4 °C,-20 °C

    Storage Comment

    • For unopened kit: All the reagents should be kept according to the labels on vials. The Standard, Detection Reagent A, Detection Reagent B and the 96-well strip plate should be stored at -20°C upon receipt while the others should be at 4°C.
    • For opened kit: When the kit is opened, the remaining reagents still need to be stored according to the above storage condition. Besides, please return the unused wells to the foil pouch containing the desiccant pack, and reseal along entire edge of zip-seal.
      Note: It is highly recommended to use the remaining reagents within 1 month provided this is within the expiration date of the kit.
    • For ELISA kit, 1 day storage at 37°C can be considered as 2 months at 4°C, which means 3 days at 37°C equaling 6 months at 4°C.

    Expiry Date

    6 months