Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

Trastuzumab Antibody ELISA Kit

This Colorimetric ELISA kit is designed for the quantitative measurement of Human Trastuzumab Antibody.
Catalog No. ABIN8091959
$643.65
Plus shipping costs $50.00
96 tests
Shipping to: United States
Delivery in 13 to 17 Business Days

Quick Overview for Trastuzumab Antibody ELISA Kit (ABIN8091959)

Target

See all Trastuzumab Antibody products
Trastuzumab Antibody

Reactivity

  • 3
  • 1
  • 1
  • 1
Human

Detection Method

Colorimetric

Method Type

Sandwich ELISA

Detection Range

0.781 ng/mL - 50 ng/mL

Application

ELISA

Sample Type

Cell Culture Supernatant, Cell Samples, Plasma, Serum, Tissue Lysate
  • Minimum Detection Limit

    0.781 ng/mL

    Purpose

    Human Anti-trastuzumab antibody ELISA Kit

    Analytical Method

    Quantitative

    Specificity

    Specifically binds with Anti-trastuzumab , no obvious cross reaction with other analogues.

    Sensitivity

    0.469 ng/mL

    Material not included

    • Microplate reader (wavelength: 450nm)
    • 37 °C incubator (CO2 incubator for cell culture is not recommenced.)
    • Automated plate washer or multi-channel pipette/5 mL pipettor
    • Precision single (0.5-10μL
    • 5-50μL
    • 20-200μL
    • 200-1000μL) and multi-channel pipette with disposable tips(Calibration is required before use.)
    • Sterile tubes and Eppendorf tubes with disposable tips
    • Absorbent paper and loading slot
    • Deionized or distilled water
  • Application Notes

    Optimal working dilution should be determined by the investigator.

    Comment

    Sandwich ELISA, Double Antigen

    Sample Volume

    100 μL

    Assay Time

    4 h

    Plate

    Pre-coated

    Protocol

    Step 1: Add 100 µL standard or sample into each well, seal the plate and statically incubate for 90 minutes at 37 °C. Washing: Wash the plate twice without immersing. Step 2: Add 100 µL biotin-antigen working solution, seal the plate and statically incubate for 60 minutes at 37 °C. Washing: Wash the plate three times and immerse for 1 min each time. Step 3: Add 100 µL HRP-Streptavidin Conjugate (SABC) working solution, seal the plate and statically incubate for 30 minutes at 37 °C. Washing: Wash the plate five times and immerse for 1 min each time. Step 4: Add 90 µL TMB substrate solution, seal the plate and statically incubate for 10-20 minutes at 37 °C. (Accurate TMB visualization control is required.) Step 5: Add 50 µL stop solution. Read at 450nm immediately and calculate.

    Restrictions

    For Research Use only
  • Handling Advice

    Avoid direct light exposure to antibody solutions

    Storage

    4 °C

    Storage Comment

    Store sealed Kit at 2-8° C. Do Not Cryopreserve. Store ELISA Microplate for 1 month at 2-8°C, Store for 12 month at -20°C. Put the leftover strips into a sealed foil bag with the desiccant

    Expiry Date

    12 months
  • Target See all Trastuzumab Antibody products

    Trastuzumab Antibody

    Alternative Name

    Anti-trastuzumab

    Target Type

    Anti-Drug Antibody

    Background

    Anti-trastuzumab antibody

    UniProt

    P04626
You are here:
Chat with us!