Phone:
+1 877 302 8632
Fax:
+1 888 205 9894 (Toll-free)
E-Mail:
orders@antibodies-online.com

IL-6 ELISA Kit

High Sensitivity IL6 Reactivity: Human Colorimetric Sandwich ELISA 0.78 pg/mL - 50 pg/mL Cell Culture Supernatant, Cell Lysate, Plasma, Serum, Tissue Homogenate
Catalog No. ABIN6999350
  • Target See all IL-6 (IL6) ELISA Kits
    IL-6 (IL6) (Interleukin 6 (IL6))
    Reactivity
    • 23
    • 18
    • 17
    • 9
    • 8
    • 8
    • 5
    • 5
    • 5
    • 4
    • 4
    • 4
    • 2
    • 2
    • 2
    • 2
    • 2
    • 1
    • 1
    • 1
    Human
    Detection Method
    Colorimetric
    Method Type
    Sandwich ELISA
    Detection Range
    0.78 pg/mL - 50 pg/mL
    Minimum Detection Limit
    0.78 pg/mL
    Application
    ELISA
    Purpose
    The kit is a high sensitivity sandwich enzyme immunoassay for in vitro quantitative measurement in various sample types.
    Sample Type
    Cell Culture Supernatant, Cell Lysate, Plasma, Serum, Tissue Homogenate
    Analytical Method
    Quantitative
    Specificity
    This assay has high sensitivity and excellent specificity for detection of Interleukin 6.
    Sensitivity
    0.35 pg/mL
    Grade
    High Sensitivity
    Components
    • Pre-coated, ready to use 96-well strip plate, flat buttom
    • Plate sealer for 96 wells
    • Reference Standard
    • Standard Diluent
    • Detection Reagent A
    • Detection Reagent B
    • Assay Diluent A
    • Assay Diluent B
    • Reagent Diluent (if Detection Reagent is lyophilized)
    • TMB Substrate
    • Stop Solution
    • Wash Buffer (30 x concentrate)
    • Instruction manual
    Top Product
    Discover our top product IL6 ELISA Kit
  • Sample Volume
    100 µL
    Assay Time
    3 h
    Plate
    Pre-coated
    Protocol
    1. Prepare all reagents, samples and standards,
    2. Add 100μL standard or sample to each well. Incubate 1 hours at 37 °C,
    3. Aspirate and add 100μL prepared Detection Reagent A. Incubate 1 hour at 37 °C,
    4. Aspirate and wash 3 times,
    5. Add 100μL prepared Detection Reagent B. Incubate 30 minutes at 37 °C,
    6. Aspirate and wash 5 times,
    7. Add 90μL Substrate Solution. Incubate 10-20 minutes at 37 °C,
    8. Add 50μL Stop Solution. Read at 450nm immediately.
    Reagent Preparation
    1. Bring all kit components and samples to room temperature (18-25 °C) before use. If the kit is not used up all at once, remove only the strips and reagents for the current experiment and leave the remaining strips and reagents in the desired condition.
    2. Standard - Reconstitute the Standard with 0.5mL of Standard Diluent, keep for 10 minutes at room temperature, shake gently (not to foam). The concentration of the standard in the stock solution is 50pg/mL. Please prepare 7 tubes containing 0.25mL of Standard Diluent and use the diluted standard to produce a double dilution series (50pg/mL, 25pg/mL, 12.5pg/mL, 6.25pg/mL, 3.12pg/mL, 1.56pg/mL, 0.78pg/mL, and the last EP tubes with Standard Diluent is the blank as 0pg/mL). Mix each tube thoroughly before the next transfer.
    3. Detection Reagent A and Detection Reagent B - Spin or centrifuge the stock of Detection Reagent A and B briefly before use. Dilute to working concentration (1:100) with Assay Diluent A or B, respectively.
    4. Wash Solution - Dilute 20 mL of Wash Solution Concentrate (30x) with 580 mL of deionized or distilled water to make 600 mL of Wash Solution (1x).
    5. TMB Substrate - Aspirate the required amount of solution with sterile tip and do not return the residual solution back into the vial.

    Note:

    1. Serial dilution directly in the wells is not recommended.
    2. Prepare standard within 15 minutes before assay. Do not dissolve the reagents directly at 37 °C.
    3. Detection Reagent A and B are sticky solutions, so pipette them slowly to reduce volume errors.
    4. Reconstitute Standard or working solutions of Detection Reagent A and B carefully according to instructions, avoiding foaming and mixing gently until crystals are completely dissolved. To minimize inaccuracy caused by pipetting, use small volumes and ensure pipettes are calibrated. It is recommended to aspirate more than 10 µL for one-time pipetting.
    5. The reconstituted Standard, Detection Reagent A and B can only be used once.
    6. When crystals have formed in the Wash Solution concentrate (30x), warm it to room temperature and mix gently until the crystals are completely dissolved.
    7. Contaminated water or preparation containers affect the detection result.
    Sample Preparation
    • It is recommended to use fresh samples without long storage, otherwise protein degradation and denaturationmay occur in these samples, leading to false results. Samples should therefore be stored for a short periodat 2 - 8 °C or aliquoted at -20 °C (≤1 month) or -80 °C (≤ 3 months). Repeated freeze-thawcycles should be avoided. Prior to assay, the frozen samples should be slowly thawed and centrifuged toremove precipitates.
    • If the sample type is not specified in the instructions, a preliminary test is necessary to determinecompatibility with the kit.
    • If a lysis buffer is used to prepare tissue homogenates or cell culture supernatant, there is a possibilityof causing a deviation due to the introduced chemical substance.The recommended dilution factor is for reference only.
    • Please estimate the concentration of the samples before performing the test. If the values are not in therange of the standard curve, the optimal sample dilution for the particular experiment has to be determined.Samples should then be diluted with PBS (pH =7.0-7.2).
    Assay Precision
    Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level of target were tested 20 times on one plate, respectively.
    Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level of target were tested on 3 different plates, 8 replicates in each plate.
    CV(%) = SD/meanX100
    Intra-Assay: CV < 10%
    Inter-Assay: CV < 12%
    Restrictions
    For Research Use only
  • Storage
    4 °C/-20 ° C
    Storage Comment
    1. For unopened kit: All reagents should be stored according to the labels on the vials. The Standard, Detection Reagent A, Detection Reagent B, and 96-well Strip Plate should be stored at -20 °C upon receipt, while the other reagents should be stored at 4 °C.
    2. For opened kits: the remaining reagents must be stored according to the above storage conditions. In addition, please return the unused wells to the foil pouch containing the desiccant and seal the foil pouch with the zipper.
    Expiry Date
    6 months
  • Target See all IL-6 (IL6) ELISA Kits
    IL-6 (IL6) (Interleukin 6 (IL6))
    Alternative Name
    Interleukin 6 (IL6 Products)
    Synonyms
    BSF2 ELISA Kit, HGF ELISA Kit, HSF ELISA Kit, IFNB2 ELISA Kit, IL-6 ELISA Kit, Il-6 ELISA Kit, ILg6 ELISA Kit, Ifnb2 ELISA Kit, il6 ELISA Kit, CHIL-6 ELISA Kit, interleukin 6 ELISA Kit, interleukin-6 ELISA Kit, IL6 ELISA Kit, Il6 ELISA Kit, il-6 ELISA Kit, IL-6 ELISA Kit
    Background
    MGI2-A, MGI2A, HGF, BSF2, HSF, IFNB2, B-Cell Stimulatory Factor-2, Hybridoma/Plasmacytoma Growth Factor, Hepatocyte Stimulating Factor, Cytotoxic T-Cell Differentiation Factor
    Pathways
    TLR Signaling, Hormone Transport, Negative Regulation of Hormone Secretion, Myometrial Relaxation and Contraction, Positive Regulation of Immune Effector Process, Production of Molecular Mediator of Immune Response, Regulation of Carbohydrate Metabolic Process, Autophagy, Cell RedoxHomeostasis, Cancer Immune Checkpoints, Inflammasome
You are here: