GAS6 ELISA Kit
Quick Overview for GAS6 ELISA Kit (ABIN8055560)
Target
See all GAS6 ELISA KitsBinding Specificity
Reactivity
Detection Method
Method Type
Detection Range
Application
Sample Type
-
-
Minimum Detection Limit
- 62.5 pg/mL
-
Purpose
- Sandwich Human GAS6 ELISA Kit to quantitate Human GAS6 in cell culture supernatants, serum and plasma (heparin or EDTA) .
-
Analytical Method
- Quantitative
-
Specificity
-
Expression system for standard: NS0
Immunogen sequence: A49-A678Capture antibody and Detection antibody: A polyclonal antibody from goat|a biotinylated detection polyclonal antibody from goat
-
Cross-Reactivity (Details)
- There is no detectable cross-reactivity with other relevant proteins.
-
Sensitivity
- < 10 pg/mL
-
Components
-
- Pre-coated 96-well strip microplate
- Standard
- Biotinylated antibody (100x)
- Avidin-Biotin-Peroxidase Complex (100x)
- Sample Diluent
- Antibody Diluent
- Avidin-Biotin-Peroxidase Diluent
- Color Developing Reagent (TMB)
- Stop Solution
- Wash Buffer (25x)
- Adhesive plate sealers
-
Material not included
- Microplate Reader capable of reading absorbance at 450nm. Incubator. Automated plate washer (optional). Pipettes and pipette tips capable of precisely dispensing 0.5 μL through 1 mL volumes of aqueous solutions. Multichannel pipettes are recommended for large amount of samples. Deionized or distilled water. 500 mL graduated cylinders. Test tubes for dilution.
-
-
-
-
Application Notes
- Before using Kit, spin tubes and bring down all components to bottom of tube. Duplicate well assay was recommended for both standard and sample testing.
-
Sample Volume
- 100 μL
-
Plate
- Pre-coated
-
Protocol
- ELISA Kit is based on standard sandwich enzyme-linked immune-sorbent assay technology. An antibody has been precoated onto 96-well plates. Standards and test samples are added to the wells, a biotinylated detection antibody specific for target is added subsequently and then followed by washing with PBS or TBS buffer. Avidin-Biotin-Peroxidase Complex was added and unbound conjugates were washed away with PBS or TBS buffer. HRP substrate TMB was used to visualize HRP enzymatic reaction. TMB was catalyzed by HRP to produce a blue color product that changed into yellow after adding acidic stop solution. The density of yellow is proportional to the target amount in sample captured in plate.
-
Assay Precision
-
Intra-Assay Precision (CV%): 4.3%, 6.5%, 4.7%
Inter-Assay Precision (CV%): 6.2%, 6.9%, 7.1% -
Restrictions
- For Research Use only
-
-
-
Handling Advice
- Avoid multiple freeze-thaw cycles.
-
Storage
- 4 °C,-20 °C
-
Storage Comment
- Store at 4°C for 6 months, at -20°C for 12 months. Avoid multiple freeze-thaw cycles (Ships with gel ice, can store for up to 3 days in room temperature. Freeze upon receiving.)
-
Expiry Date
- 12 months
-
-
- GAS6 (Growth Arrest-Specific 6 (GAS6))
-
Alternative Name
- GAS6
-
Background
-
Background: Growth arrest-specific 6, also known as GAS6, is a human gene coding for the Gas6 protein. It is mapped to 13q34. This gene encodes a gamma-carboxyglutamic acid (Gla)-containing protein thought to be involved in the stimulation of cell proliferation. This gene is frequently overexpressed in many cancers and has been implicated as an adverse prognostic marker. Elevated protein levels are additionally associated with a variety of disease states, including venous thromboembolic disease, systemic lupus erythematosus, chronic renal failure, and preeclampsia. Gas6 is a gamma-carboxyglutamic acid (Gla) domain-containing protein thought to be involved in the stimulation of cell proliferation.
Gene Full Name: growth arrest specific 6
-
Gene ID
- 2621
-
UniProt
- Q14393
-
Pathways
- RTK Signaling, Carbohydrate Homeostasis, Production of Molecular Mediator of Immune Response
Target See all GAS6 ELISA Kits
-